Image-plate sensitivity to x rays at 2 to 60 keV

  • M. J. Rosenberg
  • , D. B. Thorn
  • , N. Izumi
  • , D. Williams
  • , M. Rowland
  • , G. Torres
  • , M. Haugh
  • , P. Hillyard
  • , N. Adelman
  • , T. Schuler
  • , M. A. Barrios
  • , J. P. Holder
  • , M. B. Schneider
  • , K. B. Fournier
  • , D. K. Bradley
  • , S. P. Regan

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

The sensitivity of Fuji SR and MS image plates (IPs) used in x-ray spectrometers on OMEGA and the National Ignition Facility has been measured using two techniques. A set of radioisotopes has been used to constrain image-plate sensitivity between 6 and 60 keV, while a Manson source has been used to expose image plates to x rays at energies between 1.5 and 8 keV. These data have shown variation in sensitivity on the order of 5% for a given IP type and scanner settings. The radioisotope technique has also been used to assess IP fading properties for MS-type plates over long times. IP sensitivity as a function of scanner settings and pixel size has been systematically examined, showing variations of up to a factor of 2 depending on the IP type. Cross-calibration of IP scanners at different facilities is necessary to produce a consistent absolute sensitivity curve spanning the energy range of 2-60 keV.

Original languageEnglish
Article number013506
JournalReview of Scientific Instruments
Volume90
Issue number1
DOIs
StatePublished - Jan 1 2019
Externally publishedYes