Skip to main navigation Skip to search Skip to main content

In Situ Ion Counting for Improved Implanted Ion Error Rate and Silicon Vacancy Yield Uncertainty

Research output: Contribution to journal › Article › peer-review

32 Scopus citations

Abstract

An in situ counted ion implantation experiment improving the error on the number of ions required to form a single optically active silicon vacancy (SiV) defect in diamond 7-fold compared to timed implantation is presented. Traditional timed implantation relies on a beam current measurement followed by implantation with a preset pulse duration. It is dominated by Poisson statistics, resulting in large errors for low ion numbers. Instead, our in situ detection, measuring the ion number arriving at the substrate, results in a 2-fold improvement of the error on the ion number required to generate a single SiV compared to timed implantation. Through postimplantation analysis, the error is improved 7-fold compared to timed implantation. SiVs are detected by photoluminescence spectroscopy, and the yield of 2.98% is calculated through the photoluminescence count rate. Hanbury-Brown-Twiss interferometry is performed on locations potentially hosting single-photon emitters, confirming that 82% of the locations exhibit single photon emission statistics.
Original languageEnglish
Pages (from-to)3212-3218
Number of pages7
JournalNano Letters
Volume22
Issue number8
DOIs
StatePublished - Apr 27 2022
Externally publishedYes

Fingerprint

Dive into the research topics of 'In Situ Ion Counting for Improved Implanted Ion Error Rate and Silicon Vacancy Yield Uncertainty'. Together they form a unique fingerprint.